Our protocols for mass spectrometry have been developed to provide maximum recovery of peptides post-digestion and to ensure that interfering agents (e.g. SDS, urea, etc.) are not present. Robotic digestion also minimizes the risk of keratin contamination and allows for high throughput sample processing. Specific for MALDI analysis a preparation of matrix (a-cyano 4-hydroxy-cinnamic acid) has been developed to provide a homogeneous crystal size and maximum signal response across the entire spot area, thus minimizing the reliance on “hot spots”. We also utilize µC18 ZipTips to eliminate salts that cause a rising baseline at low m/z.
The protocol we follow for in-gel digestion (including reduction and alkylation) is as follows:
Reagents
Protocol
The final supernatant has a volume of ~20 µL. This solution is used directly for mass spectrometric analysis. We do not use (or recommend) organic solvent (typically acetonitrile) for further extraction of peptides. See acetonitrile extraction of peptides technical note.
MALDI/MS Data acquisition is performed on an Applied Biosystems Super-DE-STR instrument. We search databases using ProFound or MS-Fit. A set of typical data is presented in our sensitivity technical note.